The ESA612 Electrical Safety Analyzer represents the next generation in test equipment for biomedical and clinical engineering professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.
With selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751) and AN/NZS 3551
The versatility of the multifaceted ESA612 tester is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
Features
Key Features:
Portable, ergonomic, lightweight, and easy to use
Large, easy-to-read display with adjustable contrast
Human-factors-designed user interface
Tilt stand design for stand-up testing in field environments
Five applied parts jacks and easy ECG snap connection with optional expander box
ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
Replaceable mains fuses keep your unit in the field and out of the repair shop
Internal memory for 100 test records
20 A at 120 V current capability
USB connection for use with Ansur and Data Viewer software (for memory download to PC)
Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center)
Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
Specifications
Specifications
Voltage
Range (mains voltage)
90 V ac rms to 132 V ac rms
180 V ac rms to 264 V ac rms
Range (accessible voltage)
0 V ac rms to 300 V ac rms
Accuracy
± (2 % of reading + 0.2 V)
Voltage tests
Mains and point to point
Earth resistance
Modes
Two terminal
Test current
> 200 mA ac
Ranges
0 Ω to 2 Ω
Accuracy
± (2 % of reading + 0.015 Ω)
Resistance tests
Earth resistance and point to point
Equipment current
Mode
ac rms
Range
0 A to 20 A
Accuracy
± (5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle
15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous
Leakage current
Modes
ac + dc (true rms)
ac only
dc only
* Modes are available in all leakage tests with the exception of MAP leakages that are available only in true rms.
Patient load selection (input impedance)
AAMI ES1-1993 Fig.1
IEC 60601-1: Fig 15
Crest factor
≤ 3
Ranges
0 µA to 199.9 μA
200 µA to 1999 μA
2 mA to 10 mA
Frequency response/ Accuracy
DC to 1 kHz: ± (1 % of reading + (1 μA or 1 LSB, whichever is greater)) 1 kHz to 100 kHz: ± (2 % of reading + (1 μA or 1 LSB, whichever is greater)) 1 kHz to 5 kHz (current > 1.6 mA): ± (4 % of reading + (1 µA or 1 LSD, whichever is greater)) 100 kHz to 1 MHz: ± (5% of reading + (1 μA or 1 LSB, whichever is greater))
Note: Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are + (2.5 µA or 1 LSD, whichever is greater)
Leakage tests
Ground wire (earth)
Chassis (enclosure)
Lead to ground (patient)
Lead to lead (patient auxiliary)
Lead isolation (mains on applied part)
Direct equipment
Direct applied part
Alternative equipment
Alternative applied part
Point to point
Mains on applied part
100 % of mains
Differential leakage
Ranges
75 µ A to 199 µA
200 µA to 1999 µA
2 mA to 20 mA
Accuracy
± (10 % of reading + (2 counts or 20 uA, whichever is greater))
Insulation resistance
Ranges
0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ
Accuracy
± (2 % of reading + 0.2 MΩ)
± (7.5 % of reading + 0.2 MΩ)
Source test voltage
500 V dc
250 V dc
Insulation resistance tests
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)
ECG performance waveforms
Accuracy
± 2 %
± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration
17.63 cm x 8.38 cm x 28.45 cm (6.94 in x 3.3 in x 11.2 in)
Weight
1.6 kg (3.5 lb)
Environmental
Operating temperature
10 ºC to 40 ºC (50 ºF to 104 ºF)
Storage temperature
-20 ºC to 60 ºC (-4 ºF to 140 ºF)
Operating humidity
10 % to 90 % non-condensing
Altitude
120 V ac mains supply voltage up to 5, 000 meters
230 V ac mains supply voltage up to 2,000 meters
Warranty
Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center, otherwise standard one year warranty applies)